27 132 грн. – 43 224 грн.
Explanation of each symbol of the furnace's name:
ТУ У 28.2-30676394-001:2018
(Designation structure according to TS)
C – resistance – the type of heating;
Н – chamber – a constructive feature;
О – oxidative – workspace environment;
Л – laboratory;
58 – volume of working space in cubic dm (liters);
350 – the nominal temperature in °C.
** The prices are valid subject to the signing of the product supply agreement in the version of the Manufacturer (seller). All prices include VAT.
Description
Drying cabinet СНОЛ-58/350
СНОЛ-58/350 – The drying cabinet is designed for drying various materials and carrying out analytical work at temperatures up to 350 °C.
Product delivery time
Drying cabinet СНОЛ-58/350 will be manufactured and delivered to the customer within 5 working days from the date of order (delivery in Ukraine).
Temperature controller
The muffle furnace is equipped with an analog, microprocessor or programmable temperature controller.
An analog regulator is used if there are no high requirements for the accuracy of temperature maintenance.
The microprocessor temperature controller has a power adjustment function, which is especially important when the furnace is operating at low temperatures.
The controller allows you to adjust the heating/cooling speed and heating time.
The case
The case of the drying cabinet is made of sheet steel, powder coated on a special polymer coating line, which gives the surface an original look and durability in operation.
Powder coating is very resistant to all kinds of solvents, alkalis and acids.
Door seal
Door seal is made of special heat-resistant rubber and, together with the design of the door fastening, creates the possibility of avoiding a loose fit, and provides good thermal insulation of the working chamber.
Working chamber
Working chamber made of black steel with a special coating, or stainless steel, which allows to solve all the issues that arise for the user of our products.
The chamber is completed with regiments which can be easily removed and installed in any point of the working chamber.
Fan:
An air mixing fan is installed in the back of the working chamber, which creates a high uniformity of the thermal field inside the chamber.
Chamber heating
Chamber heating carried out from all sides, and for this purpose the original tape heater is used.
Any drying cabinet has an opening for a control thermocouple or thermometer input.
Video review of the drying cabinet
Specifications
No. p / p | Name of technical specification | Unit | Value |
---|---|---|---|
1. | Rated power, less than | kW | 2,0 |
2. | Supply voltage | V | 230 |
3. | AC frequency | Hz | 50 |
4. | Number of phases | 1 | |
5. | Rated working space temperature | °С | 350 |
6. | Automatic temperature control range | °С | 50-350 |
7. | Temperature stability in stable thermal conditions* at the measurement point, without charge, not worse | °С |
±0,5** (2,0)*** |
8. | Limits of permissible temperature deviation in the working space in the temperature range in a steady thermal regime, without charge, as the difference between Tmax and Tmin (deviation from the average value), not worse from 50 to 100 °С from 100 to 200°C from 200 to 350 °С |
°С | 3,0 (± 1,5) 6,0 (±3,0) 10,0 (±5,0) |
9. | Workspace environment | air | |
10. | Electric cabinet warm-up time to nominal temperature without charge, less than | min | 40 |
11. | Dimensions of the working chamber, not less than: width x length x height | mm | 390х380х390 |
12. | Working space size, not more than width x length x height | mm | 350х350х350 |
13. | Overall dimensions, less than: width x length x height | mm | 675х660х590 |
14. | Weight, less than | kg | 47 |
* A steady thermal regime occurs after holding for at least 60 minutes at the set temperature, without the presence of external disturbances (opening doors, changing the temperature setting, changing the mains voltage, etc.)
** for TermoPro 611 controller.
*** for TermoPro A10 controller.
Package contents
Electrical cabinet СНОЛ 58/350.
Items removed from the electrical cabinet for transportation:
– shelves – 2 pcs.
Spare parts:
– fusible insert for 1A – 1 pc.
– fusible insert for 13A – 1 pc.
Operational documentation:
– a passport for a drying cabinet.
– certificate for temperature controller.
Other models
Name | Т nom, °С | Power, kW / number of phases | Presence of a fan | Working chamber | |
---|---|---|---|---|---|
Size, mm | Material | ||||
СНОЛ 20/350 | 350 | 1 / 1 | + | 300х340х200 | steel/stainless |
СНОЛ 20/350 Vario | 350 | 1 / 1 | + | 300х340х200 | steel/stainless |
СНОЛ 24/350 | 350 | 1 / 1 | – | 300х380х200 | steel/stainless |
СНОЛ 58/350 | 350 | 2 / 1 | + | 390х380х390 | steel/stainless |
СНОЛ 67/350 | 350 | 2 / 1 | – | 390х445х390 | steel/stainless |
СНОЛ 100/350 | 350 | 4 / 1 | + | 600х380х500 | steel/stainless |
СНОЛ 120/350 | 350 | 4 / 1 | – | 600х445х500 | steel/stainless |
СНОЛ 220/350 | 350 | 6 / 3 | + | 600х620х600 | steel/stainless |
СНОЛ 250/350 | 350 | 6 / 3 | – | 600х670х630 | steel/stainless |
Warranty
The manufacturer provides a warranty for the product for 12 months. Warranty on the control unit – 18 months, on the thermoregulator series TermoPro® – 50 months.